| Model Number |
6276-1-119 |
| Medical Device Problem Codes |
Malposition of Device (2616); Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
|
| |
|
Event or Problem Description
|
|
It was reported: "cone body restoration conical to move post operation in retroversion.".
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding malposition involving a restoration modular body was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection: the restoration modular body and locking modular bolt were returned for evaluation.Visual inspection of the returned device indicates damage consistent with explantation.Bony residue is visible on the body of the stem.The stem body and bolt appear unremarkable.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to the cone body restoration conical being in retroversion.The restoration modular body and locking modular bolt were returned for evaluation.Visual inspection of the returned device indicates damage consistent with explantation.Bony residue is visible on the body of the stem.The stem body and bolt appear unremarkable.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported: "cone body restoration conical to move post operation in retroversion.".
|
| |
|