• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.(b)(4).Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a diu300 22.0 diopter intraocular lens (iol) was removed from the patient's right (od) eye after being fully inserted due to posterior capsule tear.A vitrectomy, incision enlargement, and suture were required.The replacement lens is a non-johnson and johnson lens.No further information was provided.
 
Additional Manufacturer Narrative
Section d9.Device available for evaluation? yes.Returned to manufacturer on: may 6 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the handpiece was received with the complaint lens stuck in the cartridge.Inspection of the external assembly revealed that there were no issues with the handpiece.The lens was removed from the cartridge and was cleaned; revealing that the lens was creased where it had been folded inside of the cartridge.No further issues were observed with lens.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key14389837
Report Number3012236936-2022-01247
Device Sequence Number17430716
Product Code MJP
UDI-Device Identifier05050474747760
UDI-Public(01)05050474747760(17)240811
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/11/2024
Device Model NumberDIU300
Device Catalogue NumberDIU300U220
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
Patient Weight61 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-