(b)(4).Performed a visual inspection of the returned item and found it to exhibit signs of repeated use nicked/gouged has fractured on the lateral side of the post feature.All pieces were returned.The device history records were reviewed for deviations and/or anomalies.No deviations or anomalies related to the complaint were recorded.The receiving inspection report was reviewed for deviations and/ or anomalies with no deviations/ anomalies identified.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.Investigations associated with the tibial articular surface provisional (tasp) fracture were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.In addition to the visual and optical microscopy analysis, a review of the dhf and dfmea was performed.Ongoing complaint monitoring confirms that the rate of instrument fracture is within the expected risk profile as specified in the dfmea.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|