• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIKING; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIKING; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87070
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure a viking electrode catheter lost its stiffness and bent in the patient and it was difficult to remove it from the patient.The catheter was exchanged with another of the same type.The procedure was successfully completed with no patient complications.Device has been received by boston scientific and is waiting for analysis.
 
Additional Manufacturer Narrative
The device was returned to boston scientific for analysis.Per visual inspection it was confirmed that the device has a kink in the distal section, near ring #4.Media inspection confirmed the upn and batch number of the complaint.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported that during a procedure a viking electrode catheter lost its stiffness and bent in the patient and it was difficult to remove it from the patient.The catheter was exchanged with another of the same type.The procedure was successfully completed with no patient complications.Device has been received by boston scientific and is waiting for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIKING
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key14395641
Report Number2134265-2022-05509
Device Sequence Number13942770
Product Code DRF
UDI-Device Identifier08714729878056
UDI-Public08714729878056
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K971265
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2023
Device Model Number87070
Device Catalogue Number87070
Device Lot Number0028360396
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-