| Model Number |
24686 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The 80% stenosed target lesion was located in the mildly tortuous fistula.A 035/180 starter guidewire was used to cross the lesion and a 10.0 x80, 75cm gladiator balloon catheter was advanced for dilatation.However, the wire seemed to shear and the balloon was stuck on itself.The balloon was not inflated at that point, it was just positioned and noticed it was stuck in the wire.The balloon shaft was also noted to be kinked.Both the balloon and guidewire were removed together as one unit.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a gladiator device was returned for analysis.Device was received with the customers guidewire inserted in the device.A visual examination of the returned device found that the balloon had not been inflated and was refolded.No issues were noted with the balloon material which may have potentially contributed to the complaint incident.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified a kink 380mm distal from the distal end of the strain relief and damage to the tip of the device.The investigator was unable to remove the guidewire from the device and was unable to insert a new boston scientific 0.035in (0.89 mm) guidewire due to the shaft kink and damage.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The 80% stenosed target lesion was located in the mildly tortuous fistula.A 035/180 starter guidewire was used to cross the lesion and a 10.0 x80, 75cm gladiator balloon catheter was advanced for dilatation.However, the wire seemed to shear and the balloon was stuck on itself.The balloon was not inflated at that point, it was just positioned and noticed it was stuck in the wire.The balloon shaft was also noted to be kinked.Both the balloon and guidewire were removed together as one unit.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|