| Model Number |
OM-6500 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the pull-in thread of two speedscrew implants broke prematurely.The procedure was completed with non-significant surgical delay using a back-up device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 was updated.H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection of the returned instrument shows no manufacturing abnormalities.The device was non-deployed and sutures are bundle toward the tip.Anchor is still attached.Product was out of the original packaging.No packaging returned.The devices are intended for single use.Basic functional testing was conducted.Both anchors were able to deploy once the bundled sutures were removed.Sutures green/white provided with both devices have been cut.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: 1) over tensioning the suture.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during a rm suture procedure, the pull-in thread of two speedscrew implants was torn/broken prematurely.The procedure was completed with non-significant surgical delay using a back-up device.No patient complications were reported.
|
| |
|