• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FRDM CNSTR HD 36MM T12/14 +6MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FRDM CNSTR HD 36MM T12/14 +6MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient underwent a hip revision approximately 5 years post implantation due to dislocation.During the revision the liner, head, and cup were revised.The stem remained implanted.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Source: (b)(6).Medical products: 11-107024- freedom constr.Liner +5 sz 25- 320100; unknown- unknown regenerex cup- unknown; unknown-unknown stem-unknown.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-01215.Customer has indicated that the product will not be returned to zimmer biomet for investigation, requested but not returned by hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: 103531- ti low profile screw 6.5x20mm- 374180.103534- ti low profile screw 6.5x35mm- 212290.103537- ti low profile screw 6.5x50mm- 928020.123741- 3/8-24 apical hole plug 1/cup- 684860.31-323230- 3.2mmx30mm rnglc+ acet drl bit- 449430.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No additional information on the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6 component code: mechanical (g04) - head.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRDM CNSTR HD 36MM T12/14 +6MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14396806
Report Number0001825034-2022-01216
Device Sequence Number14815638
Product Code LPH
UDI-Device Identifier00880304432499
UDI-Public(01)00880304432499(17)250424(10)851230
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K043537
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2024
Device Model NumberN/A
Device Catalogue Number14-107020
Device Lot Number851230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer05/09/2022
06/06/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date05/20/2022
06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight95 KG
-
-