Model Number SYR, NPD 0.5CC 30G 100CT5/16" 8/CS
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Syringes were not returned for evaluation.Note: manufacturer contacted pharmacist in a follow-up call on 28-apr-2022 to ensure that the initial concern was resolved - able to establish contact with pharmacist who stated the initial concern has been resolved.Pharmacist stated he was unable to contact the customer to request the syringes be returned.
 
Event or Problem Description
Consumer reported complaint for the trueplus single-use insulin syringes.Pharmacist is calling on behalf of the customer.The customer reported to the pharmacist that the needles on the 30g syringes were blunt, and that they were not able to pierce the skin or the insulin vial.Customer did not bring any of the syringes to the pharmacy, only the box that the syringes had come in.The package had not been open or damaged when received by the customer.Pharmacist was not able to confirm when the product had been dispensed to the customer.The customer feels well and did not report any symptoms.No medical attention associated with the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of 14-jun-2022: h6: updated fda¿s type, findings and conclusions codes.H10: syringes were not returned for evaluation.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using syringes from the same lot.No abnormalities were observed on retain samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGES
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key14397416
Report Number1000113657-2022-00280
Device Sequence Number12026508
Product Code FMF
UDI-Device Identifier00021292005849
UDI-Public(01)00021292005849
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K993017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Pharmacist
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/22/2024
Device Model NumberSYR, NPD 0.5CC 30G 100CT5/16" 8/CS
Device Lot NumberNP21179
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/18/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-