| Model Number |
306601 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product sample was not received for investigation.A photo was received which displayed evidence that the needle point was protruding beyond the needle guard nose.A device history review (dhr) was performed, and nonconformance data identified one in-process nonconformance; however, the nonconforming product was not included in this lot and was scrapped 100%.As the actual sample was not returned, an in-depth investigation could not be performed and root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that when removing the mechanism, it clicked but not the same sound as always and generated a little resistance to reach the end- leaving part of the needle out.There have been two cases reported, one caused a puncture and the other did not.No patient injury was reported for this case.
|
| |
|
Additional Manufacturer Narrative
|
|
While performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-08788 is no longer considered reportable, please disregard any mdr reports associated with it.
|
| |
|