• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7132
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 2.00mm x 12mm emerge balloon catheter was advanced for dilatation.However, the shaft broke upon removal from the patient.The procedure was completed with the original device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous hypotube kinks to the device.At 34.3cm from the strain relief the hypotube was separated.The separated ends were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and the balloon was tightly folded.
 
Event or Problem Description
It was reported that shaft break occurred.A 2.00mm x 12mm emerge balloon catheter was advanced for dilatation.However, the shaft broke upon removal from the patient.The procedure was completed with the original device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14398268
Report Number2134265-2022-05444
Device Sequence Number17515370
Product Code LOX
UDI-Device Identifier08714729806080
UDI-Public08714729806080
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2024
Device Model Number7132
Device Catalogue Number7132
Device Lot Number0028955854
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-