| Model Number |
7132 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.00mm x 12mm emerge balloon catheter was advanced for dilatation.However, the shaft broke upon removal from the patient.The procedure was completed with the original device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous hypotube kinks to the device.At 34.3cm from the strain relief the hypotube was separated.The separated ends were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and the balloon was tightly folded.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.00mm x 12mm emerge balloon catheter was advanced for dilatation.However, the shaft broke upon removal from the patient.The procedure was completed with the original device.There were no patient complications nor injuries reported.
|
| |
|