• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO COIL; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO COIL; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number PV-2-4-3D
Medical Device Problem Codes Separation Failure (2547); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
See manufacturer report # 2029214-2022-00820 for another coil involved in this event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that there was resistance with 4 concerto coils and the coil wouldn't detach.The patient was undergoing treatment for prostatic artery embolization (pae).The patient's blood flow and vessel tortuosity were normal.It was reported that two coils were stuck in the proximal part of the microcatheter, one was stuck in the catheter hub, and one was stuck in the distal part of the catheter. the coils had pushed out of the introducer sheath smoothly.There was no damage to the catheter or coils. the patient did not experience any injury or complications.The devices were prepared according to the instructions for use (ifu).
 
Additional Manufacturer Narrative
H3: ¿ as found condition: the concerto coil was returned for analysis within a shipping box; within primary and secondary plastic bio-pouches; within an opened concerto outer carton and within a dispenser coil.There was no instant detacher returned with the device.Information regarding the micro catheter was used in the event was not reported (model and lot number).Therefore, any contributing factors could not be assessed.¿ visual inspection/damage location details: no damages or irregularities were found with the introducer sheath.The actuator interface was securely attached to the coupler tube.No damage or irregularities were found with the actuator interface, break indicator o r the rest of the pusher.No evidence of mechanical detachment using an instant detacher or manual detachment by breaking the break indicator were found.The concerto pushwire was found to be bent within introducer sheath at ~135.0cm from proximal end.In addition, the concerto pushwire appeared to be bent at ~154.0cm, and ~155.8cm from proximal end.The concerto implant coil was still attached to the pushwire.No damage was found with the shield coil.The concerto implant coil was found damaged and stretched.The coin was found still against the lumen stop.The shield coil was found intact with the implant coil still attached.No other anomalies were observed.¿ testing/analysis: the concerto implant coil was unable to pushed out from introducer sheath, therefore the implant coil was pulled out proximally.An in-house instant detacher was then used to detach the concerto coil.The implant coil was successfully detached on the first attempt.The lumen stop inner diameter (id) and the retainer ring inner diameter (id) were measured to be within specifications.However, the concerto coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.¿ conclusion: based on the analysis performed, the customer report of ¿non-detach¿ and ¿catheter resistance/stuck in catheter hub¿ were confirmed.The concerto coil was returned with the implant coil was still attached to the pushwire.However, the root cause could not be determined as the returned axium coil was successfully detached on the first attempt using an-house instant detacher.Possible causes for coil resistance at catheter hub are, but not limited to, failure to hydrate the coil prior to use, failure to utilize continuous flush, failure to use a compatible microcatheter for delivery, and use of an improper hubbing technique (over tightening of the rhv/sheath firmly seated into hub).Improper hubbing technique could not be ruled out as a potential root cause for the event.The customer reported preparing the concerto coil prior to use (which includes implant coil hydration).It is likely that the damage to the implant coil and the pushwire occurred during the attempt to push the coil through the micro catheter against the reported resistance.However, the cause of resistance could not be determined.Since the micro catheter used in the event was not returned, any contribution of the micro catheter towards the ¿coil resistance/stuck in catheter¿ could not be assessed.As the model and lot number of the micro catheter was not reported, compatibility could not be assessed.There was no non-conformance to specification identified that led to the reported issues.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received indicating that one of the coils failed to detach.The cause of the issue was believed to be due to an issue with the catheter.The catheter was a boston scientific tru select.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO COIL
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14399445
Report Number2029214-2022-00821
Device Sequence Number14854444
Product Code KRD
UDI-Device Identifier00847536033702
UDI-Public00847536033702
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/16/2024
Device Model NumberPV-2-4-3D
Device Catalogue NumberPV-2-4-3D
Device Lot NumberB300196
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-