Model Number 0998-UC-3023-53
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The full name of the event site was shortened due to field character limit; the full name is (b)(6).Device not accessible for testing: additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
 
Event or Problem Description
It was reported that the cs300 intra-aortic balloon pump (iabp) had a fiber optic issue.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event or Problem Description
It was reported that during pre use check, the cs300 intra-aortic balloon pump (iabp) had a fiber optic issue.There was no patient involvement.
 
Additional Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse cleaned the fiber optic module.The fse then performed all functional and safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14399589
Report Number2249723-2022-01125
Device Sequence Number5415754
Product Code DSP
UDI-Device Identifier10607567108704
UDI-Public10607567108704
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-UC-3023-53
Device Catalogue Number0998-UC-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer01/26/2023
04/14/2023
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date02/07/2023
04/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2011
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-