OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 0500318E |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/29/2022 |
Event Type
malfunction
|
Event Description
|
A user facility nurse reported that a dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak that could be visually observed within the hansen connection.The blood leak occurred and was observed before reaching the blood leak detector on the machine, a fresenius 4008s machine, to trigger a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 200 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be unavailable to be returned to the manufacturer for evaluation.
|
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Manufacturer Narrative
|
Additional information: g1 plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
|
Event Description
|
A user facility nurse reported that a dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak that could be visually observed within the hansen connection.The blood leak occurred and was observed before reaching the blood leak detector on the machine, a fresenius 4008s machine, to trigger a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 200 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be unavailable to be returned to the manufacturer for evaluation.
|
|