MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Component Incompatible (1108); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx1755 showed two other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "i couldn¿t pass the guide wire through the needle with the blue arrow to it.If you look where the blue arrow is compared to the one that worked, there is a large metal piece (or something similar).That is where the guide wire would not pass at." additional info.Received 04/28/2022: was there any patient harm reported? if yes, please describe the harm, how it was treated and the outcome.
 
Event or Problem Description
It was reported "i couldn't pass the guide wire through the needle with the blue arrow to it.If you look where the blue arrow is compared to the one that worked, there is a large metal piece (or something similar).That is where the guide wire would not pass at.Additional info.Received 04/28/2022.Was there any patient harm reported? if yes, please describe the harm, how it was treated and the outcome.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: previous complaints, patient severity, applicable manufacturing records, submitted photograph, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficulty advancing the guidewire through the introducer needle due to an obstructing metal piece was inconclusive with the provided information.A photograph was provided for review which depicted two 21ga introducer needles.A blue arrow had been digitally drawn on the image and pointed towards the needle cannula within one of the introducer needles.Blood residue was seen within both needles and the needle cannula appeared similar in both.No apparent unexpected metallic piece was seen within the photographs and it was likely that the object which appeared foreign was the needle cannula.No observable evidence suggested the cause of the difficulty advancing the guidewire through the needle.A review of the manufacture records associated with the provided lot number found no potential cause identified during product manufacture.Should the physical sample be returned at a later time, this investigation will be updated.This event will be recorded for future trending and monitoring of this complaint type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14399972
Report Number3006260740-2022-01784
Device Sequence Number17505791
Product Code LJS
UDI-Device Identifier00801741154850
UDI-Public(01)00801741154850
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K051672
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1275108FD2
Device Lot NumberREFX1755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient SexMale
Patient Weight98 KG
-
-