• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00558940
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device is not available for return; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre rx biliary dilatation balloon was used in the stomach and pancreas during a dilatation of hotaxios procedure performed on (b)(6) 2022.During the procedure, it was noticed that the dark part of the balloon catheter, the rx tunnel, stayed inside the working channel when the physician attempted to remove the balloon.The dilatation was completed at this time.The endoscope was removed and the nurse took out the black part using a cleaning brush and the procedure was completed with a therapeutic gastroscope.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14400411
Report Number3005099803-2022-02596
Device Sequence Number20202238
Product Code FGE
UDI-Device Identifier08714729965497
UDI-Public08714729965497
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K172520
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2024
Device Model NumberM00558940
Device Catalogue Number5894
Device Lot Number0028521635
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2022
Initial Report FDA Received Date05/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-