• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT TSH REAGENT KIT; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT TSH REAGENT KIT; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number 7K62-35
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely decreased architect tsh results for 2 samples when running on the architect i2000sr analyzer.The following data was provided: (b)(6) 2022 sid(b)(6): initial tsh: 2.15uiu/ml, previous result: 4.19 uiu/ml.(b)(6) 2022 sid (b)(6): initial tsh: 0.46 uiu/ml, previous result: 1.09 uiu/ml.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
Patient identifier: sid (b)(6) and sid (b)(6) 2 different patients.All available patient information is included.Additional patient details are not available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The complaint investigation for falsely depressed architect tsh results included a search for similar complaints, and the review of complaint text, trending data, labeling, and device history records.Return testing was not performed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.A ticket search by lot indicates that the reagent lot performs as expected for the product.A review of tracking and trending determined no related trends identified.Customer field data was used to assess the performance of the architect tsh assay using worldwide data.The median patient population result for lot 32427ud00 is comparable with all other lots in the field and falls within established baselines, confirming no systemic issue for the lot.The device history record was reviewed and did not identify any non-conformances or deviations associated with reagent lot 32427ud00 and complaint issue.Labeling was reviewed and adequately addresses the issue under review.Based on the investigation, no systemic issue or product deficiency for architect tsh reagent lot 32427ud00 was identified.Section d3 suspect medical device was updated to include new mfg site email section g1 all manufacturers was updated to include new mfg site email.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TSH REAGENT KIT
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14400449
Report Number3005094123-2022-00103
Device Sequence Number12125396
Product Code JLW
UDI-Device Identifier00380740105969
UDI-Public00380740105969
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K983442
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2023
Device Model Number7K62-35
Device Catalogue Number07K62-35
Device Lot Number32427UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2K PROC MOD, 03M74-01, (B)(6) ; ARC I2K PROC MOD, 03M74-01, (B)(6)
Patient SexUnknown
-
-