• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367844
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 7.2mg blood collection tubes it fail to fill the tube to the indicated line.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: ¿tubes do not exhaust the vacuum and fail to fill the tube to the indicated line and even others, it looks like they have no vacuum.¿.
 
Additional Manufacturer Narrative
H.6.Investigation: bd received 1 photo from the customer in support of this complaint.The photo shows the tube with no specimen in the tube, it is unknown if the tube was used from the photo.Therefore, 10 production lot in-house retention tubes from the bd inventory were visually inspected and functionally tested and, the customer' s indicated failure mode of underfill was not observed as all tubes were within specification limits.Bd was unable to confirm the customer¿s indicated failure mode as no samples were received, and the defect was not observed in the photo or the retention sample testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.The quantity of blood drawn into evaluated tubes varies with altitude, ambient temperature, barometric pressure, tube age, venous pressure, and filling technique.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 7.2mg blood collection tubes it fail to fill the tube to the indicated line.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: ¿tubes do not exhaust the vacuum and fail to fill the tube to the indicated line and even others, it looks like they have no vacuum.¿.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 12-jul-2022 h6: investigation summary bd received 2 samples and 1 photo from the customer in support of this complaint.The photo was evaluated and shows a tube with no specimen in the tube.It is unknown if the tube was used from the photo.The 2 sample tubes were functionally tested and the issue of underfill was not observed as all tubes were within specification limits.Additionally, 10 retention tubes from the bd inventory were functionally tested with no issues being observed.Bd was unable to confirm and/or duplicate the customer¿s indicated failure mode because the defect was not observed in the sample, photo, and retention testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 7.2mg blood collection tubes it fail to fill the tube to the indicated line.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: ¿tubes do not exhaust the vacuum and fail to fill the tube to the indicated line and even others, it looks like they have no vacuum.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14400686
Report Number1917413-2022-00263
Device Sequence Number5490001
Product Code JKA
UDI-Device Identifier50382903678445
UDI-Public50382903678445
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2022
Device Model Number367844
Device Catalogue Number367844
Device Lot Number1195783
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer05/25/2022
07/19/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date05/27/2022
08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-