| Model Number |
4540008-02 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint is still under investigation.A follow-up report will be provided, as soon as investigation has been completed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility / translation of user facility information by bbm sales organization in france: "pump-flow rate-fast".According to the customer: "too fast diffusion while the duration is supposed to be in 12h with 120 ml.The nurse puts 150ml and the infusions always end before 12 hours.Consequences: potential overdose and potential antibiotic residual problem.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report # (b)(4).As no sample and no meaningful picture was provided, proper investigation on malfunction could not be performed.Hence, the complaint is considered as not confirmed.Device history record (dhr): reviewed the dhr for batch 21l11ge221, there is no abnormality and no such defect detected at in process and at final control inspection.
|
| |
|