AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCB00
Medical Device Problem Codes Break (1069); Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Unknown/asked but unavailable.Date implanted: not applicable, as the iol was not implanted.Date explanted: not applicable, as the iol was not implanted.Therefore; it was not explanted.The device was not returned for evaluation.Therefore; a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported the intraocular lens (iol) was partially inserted in the patient¿s ocular dexter (right eye).The iol was removed due to a bent haptic.The same procedure completed successfully using another lens with the same model and diopter size.There was no medical or surgical intervention required.Patient outcome was reported as going good.No further information is available.
 
Additional Manufacturer Narrative
Corrected data: upon further review, it was noted that in the initial report section h6 medical device problem code: 2339 - inaccurate delivery was captured.Device code 2339.Do not apply as the partially inserted lens was due to the haptic being bent.Therefore, this supplemental filing is to correct the medical device problem code: 2339 that was incorrectly reported.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d9: device available for evaluation: yes.Section d9: date returned to manufacturer: 07-jun-2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received coated in viscoelastic residue and particles from the folding carton.The lens was cleaned, and the haptics were observed to not be twisted; however, a haptic became detached during handling.Complaint issue (dc-haptic damaged) was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14401743
Report Number3012236936-2022-01240
Device Sequence Number5301509
Product Code HQL
UDI-Device Identifier05050474531697
UDI-Public(01)05050474531697(17)250812
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2025
Device Model NumberZCB00
Device Catalogue NumberZCB0000190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer06/20/2022
08/04/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date07/13/2022
08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-