• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC SCEPTER C OCCLUSION BALLOON;(4X10 MM); OCCLUSION BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC SCEPTER C OCCLUSION BALLOON;(4X10 MM); OCCLUSION BALLOON CATHETER Back to Search Results
Model Number BC0410C-ID-700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformance's associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for analysis but has not yet been returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.If the device is received, the investigation will be performed, and a supplemental report will be submitted.The instruction for use (ifu) include the warnings: do not exceed the maximum recommended inflation volume as balloon rupture may occur.Excessive pressure higher than 300 psi (2069kpa) may cause leakage or rupture of the balloon catheter working lumen.
 
Event or Problem Description
It was reported that during the procedure, a scepter c balloon ruptured during inflation.There was no reported intervention or injury to the patient.
 
Additional Manufacturer Narrative
Additional information: d9 (date device received).Additional information was received indicating the balloon leak was a slow one.No adverse patient effects from this complaint.Summary device evaluation: the inspection of the returned scepter c found a pinhole when inflated near the proximal bond of the balloon.The investigation could not determine what likely caused the pinhole as no other contributing factors or conditions were found during this evaluation; a review of the manufacturing records found that this lot met all release criteria from the manufacturing procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCEPTER C OCCLUSION BALLOON;(4X10 MM)
Common Device Name
OCCLUSION BALLOON CATHETER
Manufacturer (Section D)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise dr.
aliso viejo, CA 92656
7142478000
MDR Report Key14401772
Report Number2032493-2022-00197
Device Sequence Number17515396
Product Code MJN
UDI-Device Identifier00842429103388
UDI-Public(01)00842429103388(11)210831(17)240731(10)0000070322
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K110741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberBC0410C-ID-700
Device Catalogue NumberBC0410C
Device Lot Number0000070322
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-