EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 3000
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported through implant patient registry that a patient with a 19 mm 3000 aortic valve implanted in the mitral position for 4 years and 2 months underwent a valve-in-valve procedure due to deterioration.The procedure was performed with a 23 mm 9750tfx transcatheter valve with no complications.At the end of the procedure the patient was hemodynamically stable and was transferred to the cvicu.Post procedure patient in recovery.Patient was discharge on pod#1.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional customer complaint.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Through further investigation, it was determined that a definitive root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
1 edwards way
ant 6.1
irvine, CA 92614
9492502731
MDR Report Key14402586
Report Number2015691-2022-05708
Device Sequence Number14825236
Product Code DYE
UDI-Device Identifier00690103162233
UDI-Public(01)00690103162233(17)181129
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2018
Device Model Number3000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-