Model Number NGP680301
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Codes Eye Injury (1845); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Information unknown/ not provided.Per regulation eu (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Telephone number: (b)(6).(b)(4).Device evaluation: as a result of this complaint, a field service engineer performed a check of the system.The vacuum values on surgeon mode 2 were checked and as per checklist and vacuum were within specifications.All values relating to the whitestar device were also within specifications, no adjustments or calibrations required.Manufacturing record evaluation: the manufacturing records for the whitestar signature pro system were reviewed.The product was manufactured and released according to specification.A review of the device history record, complaint trending, and risk documentation for this device was performed.There were no discrepancies or non-conformances experienced during manufacturing.The system and its components met all specifications prior to being released.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.The device functioned as intended.Attempts have been made to obtain missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
The customer reported that they had problems with occlusion.When switching to mode 2 while doing phacoemulsification on the soft cataract, the machine was not able to suck up the cataract.It kept getting pushed away from the phaco tip.The customer had to switch to ia (irrigation and aspiration) mode and use the hand piece to finish the cataract suction process.They also needed to do a vitrectomy cut due to injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE PRO
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14402616
Report Number3012236936-2022-01243
Device Sequence Number5452000
Product Code HQC
UDI-Device Identifier05050474596146
UDI-Public(01)05050474596146
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K151636
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNGP680301
Device Catalogue NumberNGP680301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Initial Report FDA Received Date05/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-