| Model Number |
TMICL12.6 |
| Medical Device Problem Codes |
Misfocusing (1401); Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter states an implantable collamer lens was implanted into the patient's right (od) eye.The surgeon reports an oversized icl.Attempts to obtain additional information have not been successful.
|
| |
|
Additional Manufacturer Narrative
|
|
B1 - corrected to adverse event in initial mdr.B2 - corrected to required intervention to prevent impairment/damage (devices) in initial mdr.B3 - date of event (b)(6) 2022 should be added to the initial mdr.B5 - "the reporter indicated the surgeon implanted a 12.6mm tmicl 12.6 implantable collamer lens, -17.00/2.0/089 (sphere/cylinder/axis), in the patients right eye (od) on (b)(6) 2022.The lens was explanted on (b)(6) 2022 due low vault with rotation, lens "dislocation"/subluxation, and blurred vision.The lens was replaced with another same model/ longer length lens and the problem was resolved.The reporter marked the cause of the event as the device." should be added to the initial mdr.D6a - implant date (b)(6) 2022 should be added to initial mdr.D6b - explant date (b)(6) 2022 should be added to initial mdr.H1 - corrected to serious injury in initial mdr.H6 - health effect - impact code: 4627 should be added to initial mdr.H6 - health effect - clinical code: 2137 should be added to initial mdr.H6 - health effect - device code: 1401 -2923 should be added to initial mdr.H6 - health effect - type if investigation code: 4110-41111 :" lens work order search - no similar complaints reported for units within the same lot." should be added to initial mdr.Claim #: (b)(4).
|
| |
|
Search Alerts/Recalls
|