| Model Number |
441386 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ fx, instrument bottom, packaged false positive results were acquired.There was no patient impact.The following information was provided by the initial reporter: "it was reported by the customer that drawer c had false positives.2 false positives.What confirmatory testing was performed that determined the results were erroneous? sub culture, no confirmed positives.Were any erroneous results reported to the doctors? no.If yes, were any patients treated based on erroneous results? no.If yes, did the erroneous treatment have any adverse impact to the patient(s)? no.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation of the complaint, it has been determined that the previously submitted mfr report# 1119779-2022-00693 was sent in error.The false positive results are related to the reagent, this malfunction will be captured with a new mfr report that has yet to be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ fx, instrument bottom, packaged false positive results were acquired.There was no patient impact.The following information was provided by the initial reporter: "it was reported by the customer that drawer c had false positives 2 false positives -what confirmatory testing was performed that determined the results were erroneous? sub culture, no confirmed positives.-were any erroneous results reported to the doctors? no -if yes, were any patients treated based on erroneous results? no -if yes, did the erroneous treatment have any adverse impact to the patient(s)? no".
|
| |
|
Search Alerts/Recalls
|