| Model Number |
166816 |
| Medical Device Problem Codes |
Material Invagination (1336); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
12/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter had an unknown issue.Additional information has been requested.Per follow-up information received from (b)(6) team on 20apr2022, the foley catheter had a material deformation problem and mechanical jam problem.It was also stated that the patient experienced hemorrhage/bleeding.It was unknown if the device contributed to the hemorrhage/bleeding at this time and medical intervention was unknown.No other information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "incorrect balloon design (balloon wall thickness excessive)".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." correction: f, h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter had an unknown issue.Per follow-up information received on (b)(6) 2022, the foley catheter had a material deformation problem and mechanical jam problem.It was also stated that the patient experienced hemorrhage/bleeding.Per follow-up information received on (b)(6) 2022, foley catheter was removed and got stuck on the way out causing urethral trauma and bleeding with removal and post removal.Upon inspection of the catheter post removal, there were wrinkles and rolls in the area where the balloon was inflated, frank blood noted on the area and in the tip of the catheter.The patient also experienced urethra tissue breakdown while the catheter was inserted, despite the securement of the catheter to the patient's thigh, possibly due to the rigid nature of the product.It was unknown what medical intervention was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be "inappropriate package design".It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.A labeling review was not performed because labelling could not have prevented the reported failure.Corrections: b,g h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter had an unknown issue.Additional information has been requested.Per follow-up information received on 20apr2022, the foley catheter had a material deformation problem and mechanical jam problem.It was also stated that the patient experienced hemorrhage/bleeding.No medical intervention was reported.
|
| |
|