| Model Number |
TMICL 12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight-race:unk.Claim #: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm tmicl 12.6 implantable collamer lens of -12.0/1.5/111(sphere/cylinder/axis) was implanted into the patient's left eye (os) on (b)(6) 2022.Excessive vault and elevated iop was observed.On (b)(6) 2022 the lens was removed.A replacement lens was later implanted with a spherical model and shorter length.The replacement lens did not resolve the problem.See mfr # 2023826-2022-01553 for replacement lens.
|
| |
|
Additional Manufacturer Narrative
|
|
H6-type of investigation code : 10 in previous mdr should be corrected to 4110.Additional information: h6- device evaluation: lens returned in a micro-centrifuge vial in liquid.Visual inspection found haptic torn.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|