• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE DISTAL ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE DISTAL ATTACHMENT Back to Search Results
Model Number D-201-10704
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, oil stain adheres to the inside of the cap, and the cap was slightly crushed and deformed, so the inner diameter was out of the standard value.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
The customer reported to olympus, that during an endoscopic retrograde cholangiopancreatography (ercp) procedure, the disposable distal attachment cap fell off inside the body of the patient.It was reported that the cap was attached to the scope and the hood originally entered the operation part for couple of millimeters but was in a state of being pushed back.The attachment was immediately collected, and the procedure was completed with the same device.There were no reports of patient harm associated with this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is possible that the reported event occurred due to the effect of lubricant adhered inside of the subject attachment device.A definitive root cause cannot be identified.This issue is covered in the instructions for use: ¿ do not use vaseline (or any lubricant that contains petroleum jelly), olive oil, alcohol, or the xylocain spray to lubricate the instrument.Doing so could cause equipment damage, or cause the instrument to fall off of the endoscope inside the patient.¿ be sure to wipe away any excess lubricant before mounting the instrument, and secure the instrument firmly using medical tape with sufficient elasticity.If the instrument is not firmly secured, it could come off inside the body cavity during use and cause patient injury, such as mucous membrane damage.·if you encounter strong resistance when mounting the instrument on the endoscope, apply a water-soluble lubricant to the endoscope attachment section.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE DISTAL ATTACHMENT
Common Device Name
DISPOSABLE DISTAL ATTACHMENT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14403658
Report Number8010047-2022-08268
Device Sequence Number17511341
Product Code FDS
UDI-Device Identifier04953170278266
UDI-Public04953170278266
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K984358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD-201-10704
Device Lot Number1YK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date05/14/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-