C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1718500
Medical Device Problem Codes Difficult to Flush (1251); Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that approximately three years and eleven months post port placement procedure, the port was allegedly unable to be flushed.It was further reported that the catheter was allegedly broken.Reportedly, had procedure to remove the port and only two portions of the catheter.Snare was used to remove floating/dislocated portion of the catheter.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.(expiration date: 08/2021).
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerport ti l/p implantable port attached to a groshong catheter in two segments were returned for evaluation and two electronic photos were provided for review.The investigation is confirmed for the reported fracture, material separation and identified deformation and wear issues as a complete circumferential break was noted on the returned catheter.The edges of the complete circumferential break was noted to be rounded in one region and granular in another.The surface was noted to be granular and glossy.Further a partial compound break was noted at the proximal end of the distal catheter segment.The edges of the partial compound break on the distal catheter segment was noted to be jagged.The surface was noted to be round in one region and granular in another region.However, the investigation is inconclusive for the reported difficult to flush issue as the exact circumstances at the time of the reported event cannot be verified.The identified break is typical of flexural fatigue, which is due to the repetitive, cyclic kinking of the catheter.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 08/2021), g3, h6 (device, method).H11: h6 (result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that approximately three years eleven months post port placement procedure, the port was allegedly unable to be flushed.It was further reported that the catheter was allegedly broken.Reportedly, had procedure to remove the port and only two portions of the catheter.Snare used to remove floating/dislocated portion of the catheter.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14403818
Report Number3006260740-2022-01806
Device Sequence Number17544524
Product Code LJT
UDI-Device Identifier00801741026812
UDI-Public(01)00801741026812
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081311
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1718500
Device Catalogue Number1718500
Device Lot NumberREBQ0194
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/06/2022
Initial Report FDA Received Date05/15/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-