Model Number NBD10
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that few days post drainage catheter placement procedure, the catheter hub was allegedly found broke.It was further reported that the broken hub was removed.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that few days post drainage catheter placement procedure, the catheter hub was allegedly found broke.It was further reported that the broken hub was removed.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE BILIARY DRAIN
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14403849
Report Number3006260740-2022-01809
Device Sequence Number17530825
Product Code FGE
UDI-Device Identifier00801741036842
UDI-Public(01)00801741036842
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K955110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNBD10
Device Catalogue NumberNBD10
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer06/13/2022
Initial Report FDA Received Date05/15/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-