MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
|
Back to Search Results |
|
| Model Number |
ETLW1613C156E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: esbf3214c103ee, serial/lot #: (b)(4), ubd: 24-mar-2017, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted during the endovascular treatment of an abdominal aortic aneurysm.It was reported that during a follow up ct, it was found that the aneurysm sac increased to 65mm.A type ii endoleak was identified and a type iii endoleak was suspected.It was noted that the type iii did not appear to be a separation endoleak, but a possible hole or tear in graft.Intervention was performed. an angiogram was performed in ap view after inflating a balloon in the contra limb just below the flow divider.It demonstrated late filing of the aaa sac, confirming the type ii endoleak from the patent lumbar artery.However, when rotating the image intensifier in a more lateral projection, it was found a blush of contrast filled briskly into the sac 30-40mm distal to the gate in the 16x13x156 limb.The endoleak was just distal to gate at the flow divider at the top of the limb.It was then decided to implant a etlw1613c156e limb to reline landing above the hypogastric artery.The physician obtained access to the lumbar arteries from the liia using a microcatheter and then non mdt coils were placed to occlude the endoleak.A final angiogram was taken in the same projection which showed no flow into the sac.Per physician, the cause of the event was due to patent type ii endoleak from 2 sets of lumbar arteries with potential type iii en doleak, from the left limb at the level of the flow divider. no additional clinical sequelae were reported and the patient if fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Film analysis the reported endoleak was confirmed on the films provided; however, the type and cause of the event could not be fully determined.The aneurysm enlargement could not be confirmed.The anatomy at implant is unknown and earlier post-implant films were not provided for comparison.It is most likely that the contrast leakage observed may have been a type ii endoleak from the patent lumbar arteries.A type iii fabric endoleak cannot be ruled out and may have also been present, as the calcification observed in the area of the leak could be a factor in the occurrence on this type of endoleak.Potential aneurysm morphology changes during the implant duration may have also exacerbated fabric abrasion wear.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|