The customer reported a false negative reaction of a patient sample with cell #1 of ih-cell i-ii.The customer stated that the patient sample was previously frozen and was thawed for a validation study.The specificity of the missed antibody was anti-c (rh2).The customer tested manually and also used the ih-reader 24.Furthermore, the customer reported that the patient sample showed 1+ positive reactions with ih-panel 11 and the reagent red blood cells (rrbcs) from ortho.The customer did neither provide a product sample for investigational testing nor the patient sample that caused the false negative reaction.Therefore, our quality control laboratory tested their retention sample of the supposedly defective product lot.It was tested manually with four different anti-c (rh2) and two donor samples.All positive and negative reactions were correct.Additionally, cell #1 of ih-cell i-ii was tested for potency.The testing was done in parallel with the c-antigen positive rrbcs #1, #3 and #5 of ih-panel 11 and rrbc #1 of the current lot ih-cell i-ii.All reagent red blood cells showed comparable results.We did not observe any weakening of the c (rh2) antigen of the affected rrbc #1 of ih-cell i-ii.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Data files of the affected ih-reader were provided.A first investigation showed that the data from (b)(6) 2022, were not available, but only hard copies of the test-in-question.The investigation of the data files from (b)(6) 2022, is still ongoing.
|
|
The customer reported a false negative reaction of a patient sample with cell #1 of ih-cell i-ii.The customer stated that the patient sample was previously frozen and was thawed for a validation study.The specificity of the missed antibody was anti-c (rh2).The customer tested manually and also used the ih-reader 24.Furthermore, the customer reported that the patient sample showed 1+ positive reactions with ih-panel 11 and the reagent red blood cells (rrbcs) from ortho.Data files of the affected ih-reader were provided.A first investigation showed that the data from april 14, 2022, were not available, but only hard copies of the test-inquestion.The hard copies showed a negative reaction with cell #1 of ih-cell i-ii.The data files from april 25, 2022 were investigated.Based on the result interpretations and analyses, we could confirm that the reading module and the camera are correctly adjusted.We recognized that the pipetted volume was too little, only round about 50¿l instead of 75 l, so we highly recommended checking the pipetting step.According to our investigation we highly suspected that we were facing a low-titer antibody and the comparison with the ortho system just showed the borderline of our gel cards when testing this specific sample.The customer did not provide the allegedly defective product sample ih-cell i-ii for investigational testing, but the patient sample that had caused the false negative reaction.While we were waiting for the patient sample to arrive on our premises, our quality control laboratory tested manually their retention sample with four different anti-c (rh2) and two donor samples.All positive and negative reactions were correct.Additionally, cell #1 of ih-cell i-ii was tested for potency.The testing was conducted in parallel with the c-antigen positive rrbcs #1, #3 and #5 of ih-panel 11 and rrbc #1 of the current lot ih-cell i-ii (lot # 9214011).All reagent red blood cells showed comparable results.We did not observe any weakening of the c (rh2) antigen of the affected rrbc #1 of ih-cell i-ii.When we received the patient sample, our quality control laboratory tested it with the complained lot and also with the current lot of ih-cell.Due to the small amount of plasma the patient sample was tested manually with selected reagent red blood cells (rrbc): cell 1 of ih-cell i-ii in each case lot #9210011 and lot #9218011 with a negative result: cell #1 and #3 of ih-panel 11 in each case lot #9210011 and lot #9218011 with a result: 1+ positive with lot #9210011 and negative with #9218011.All tests were performed in parallel with a known anti-c.This anti-c showed positive reactions with all used rrbcs.Further tests, e.G.Potency testing of the patient sample was not possible due to the small quantity provided.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.Based on the investigation the complaint was classified as confirmed - epp (expected product performance).The customer did not provide the complaint sample, but a low quantity of the patient sample that caused the false negative.The testing of the retention sample confirmed the correct function of the supposedly defective lot.Nevertheless, the patient sample showed negative reactions with cell #1 of ih-cell i-ii.But based on the provided information that the patient sample showed only 1+ positive reactions in the other tests we most probably were facing a low-titer antibody.The instruction for use contained the appropriate note in the section limitation: "negative reactions will be obtained if the sample contains antibodies present in concentrations too low to be detected by the test method employed.No test method is capable of detecting all red blood cell antibodies." furthermore, we highly recommended checking the pipetting step at customer.Because the investigation showed that the pipetted volume of plasma seemed to be too low, less than 25 l.
|