• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH CELON ELITE ESG-200, E; HF-GENERATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH CELON ELITE ESG-200, E; HF-GENERATORS Back to Search Results
Model Number WA90001A
Medical Device Problem Codes Output Problem (3005); Activation Problem (4042)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Olympus was informed that during a therapeutic procedure to treat hypertrophic turbinates, the hf-generator esg-200 could not be activated.The patient had already being anesthetized when the device failure was noticed resulting in the procedure being cancelled as the issue could not be resolved.No further information was provided.
 
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for investigation/evaluation but to a regional olympus repair center in ireland.The evaluation confirmed the reported output failure and traced it back to a defective motherboard.Thus, the reported incident can be attributed to component failure.A manufacturing and quality control review was performed for the affected serial number of the hf generator without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELON ELITE ESG-200, E
Common Device Name
HF-GENERATORS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
rheinstrasse 8
teltow
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key14407000
Report Number9610773-2022-00176
Device Sequence Number20231837
Product Code GEI
UDI-Device Identifier04042761082952
UDI-Public04042761082952
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K160639
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA90001A
Device Catalogue NumberWA90001A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date05/16/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/25/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-