• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC BIS/IS-15; PACEMAKER LEAD ADAPTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC BIS/IS-15; PACEMAKER LEAD ADAPTOR Back to Search Results
Model Number BIS/IS-15
Medical Device Problem Codes Failure to Capture (1081); Device Sensing Problem (2917); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Our investigation is still in progress, follow up report will be submitted if we find any further additional information.
 
Event or Problem Description
It was reported that there was no capture at maximum output (7.5v @ 1.5ms) and the only sensed r waves were from the external pacemaker.During the implantable icd interrogation, we turned off the external pacer in order to ascertain the underlying rhythm, and it was discovered the patient's rhythm was complete heart block.The doctor then turned the external pacer back on to maintain capture.Subsequently, he successfully inserted a temporary transvenous wire and connected it to the external temporary pacemaker, where constant capture was taking place.Patient outcome is stable.No additional information is available.
 
Additional Manufacturer Narrative
The following sections were updated in follow-up 1: b1,b2, b4, g1, g3, g6, h2, h6, h10 and h11.Correction : b1,b2,h6 :there was no issue with bis/is-15 adaptor reported.The epicardial lead was not capturing and the patient was very symptomatic.Thus, b1,b2, h6 updated accordingly.There was no issue with bis/is-15 adaptor reported.The adaptor was not returned and remains implanted since the event date 3/15/2022.The investigation was focused on a review of product documentation.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections no further investigation is required.Based on the investigation, a capa is not required.The event will be re-evaluated if additional information becomes available.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS/IS-15
Common Device Name
PACEMAKER LEAD ADAPTOR
Manufacturer (Section D)
OSCOR INC
3816 desoto blvd
palm harbor FL 34683 1816
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd
palm harbor FL 34683
Manufacturer Contact
doug myers
3816 desoto blvd
palm harbor, FL 34683
7279372511
MDR Report Key14407179
Report Number1035166-2022-00065
Device Sequence Number5564327
Product Code DTD
UDI-Device Identifier00836559002680
UDI-Public00836559002680
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K925168
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2024
Device Model NumberBIS/IS-15
Device Catalogue NumberBIS/IS-15
Device Lot NumberC4-15262
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date05/16/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-