| Model Number |
30MM CENTRAL SCREW, MODULAR |
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported through the surgical outcome system that a revision surgery is required due to a prosthetic joint implant failure.The date of the primary procedure was (b)(6) 2020 for a supscap repair shoulder arthroplasty.Update: (b)(6) 2022: the central screw and one peripheral screw had broken and a revision surgery was performed on (b)(6) 2022 due to the loosening of the baseplate.The problem was identified on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Upon visual inspection, it was noted that the device was received assembled to an ar-9560-28-4.It was noted that the central screw had broken.The cause of this is undetermined.Upon functional testing, it was noted that the devices could not be separated.
|
| |
|
Search Alerts/Recalls
|