Conclusion: the reported event was confirmed.Medtronic received information that prior to use of an affinity nt oxygenator and card iotomy venous reservoir, the customer reported that the sampling port of the device was broken.Photos provided by the customer confirmed the report.Production information was reviewed, and the product had passed all in-process quality checks; manufacturing controls are in place for the inspection of each device to ensure that product meets specification prior to the release from the manufacturing facility.Further review indicated that the process of adhering the connector to the tube could have contributed to the reported event, so the most likely cause is manufacturing related.Trends for issues with this product are reviewed.B5 additional information: medtronic received additional information that since this case had taken place during the pre-operative inspection, the customer just used a different product than the broken one.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|