MAUDE Adverse Event Report: ENCORE MEDICAL L.P EMPOWR VVC KNEE; EMPOWR VVC KNEETM TIBIAL INSERT, SIZE 10, 14MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P EMPOWR VVC KNEE; EMPOWR VVC KNEETM TIBIAL INSERT, SIZE 10, 14MM Back to Search Results
Model Number 346-14-710
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient was experiencing some laxity and a poly swap was performed.The surgeon up sized the poly.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as laxity and poly swap.The previous surgery and the surgery detailed in this event occurred 1 year and 6 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to laxity and poly swap.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Additional Manufacturer Narrative
1644408-2022-00654 was reassessed and determined to be non-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR VVC KNEE
Common Device Name
EMPOWR VVC KNEETM TIBIAL INSERT, SIZE 10, 14MM
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14408431
Report Number1644408-2022-00654
Device Sequence Number11382378
Product Code JWH
UDI-Device Identifier00190446219019
UDI-Public(01)00190446219019
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180930
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/22/2024
Device Model Number346-14-710
Device Catalogue Number346-14-710
Device Lot Number452V1006
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer05/16/2022
05/16/2022
Initial Report FDA Received Date05/16/2022
Supplement Report FDA Received Date05/25/2022
06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-