Model Number 530-10-048
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection and loosening.The previous surgery and the surgery detailed in this event occurred 6 months and 2 weeks apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were kept at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr # (b)(4) associated with the main part #530-10-048, altivate reverse, humeral stem.Standard shell, sz 10x48mm which documents that out of (b)(4) parts lot, 2 parts were rejected and scraped due to low spot in porous coating, looks discolored.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection and loosening.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - the stem was loosen and possible infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE REVERSE SHOULDER
Common Device Name
ALTIVATE REVERSE, HUMERAL STEM. STANDARD SHELL, SZ 10X48MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14408436
Report Number1644408-2022-00645
Device Sequence Number16763900
Product Code KWS
UDI-Device Identifier00190446334903
UDI-Public(01)00190446334903
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190290
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/06/2025
Device Model Number530-10-048
Device Catalogue Number530-10-048
Device Lot Number1485A1056
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Initial Report FDA Received Date05/16/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-36-101 LOT 869C3254; 509-01-436 LOT 390P1072; 510-08-000 LOT 142G1396
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
-
-