• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CELESTICA ELECTRONICS S PTE LTD BIS; ELECTRODE, CUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CELESTICA ELECTRONICS S PTE LTD BIS; ELECTRODE, CUTANEOUS Back to Search Results
Model Number 186-0106
Device Problems Use of Device Problem (1670); High Readings (2459)
Patient Problems Device Overstimulation of Tissue (1991); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Event Description
According to the reporter, during intervention, the monitor displayed wrong measurement.The sensor showed a flat curve (eeg) which did not correlate with the given value of 90, which means that the patient was awake.The patient was given additional hypnotics, but the values did not change.After changing the sensor, the value dropped to 6, which means that the patient was in deep sleep.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during intervention, the monitor displayed wrong measurement.The sensor showed a flat curve (eeg) which did not correlate with the given value of 90 which means that the patient was awake.The patient was given additional hypnotics, but the values did not change.After changing the sensor, the value dropped to 6 which means that the patient was in deep sleep.The patient was not injured because of the additional hypnotics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Type of Device
ELECTRODE, CUTANEOUS
Manufacturer (Section D)
CELESTICA ELECTRONICS S PTE LTD
no. 6 serangoon north avenue 5
singapore 55491 0
SN  554910
Manufacturer (Section G)
CELESTICA ELECTRONICS S PTE LTD
no. 6 serangoon north avenue 5
singapore 55491 0
SN   554910
Manufacturer Contact
clay chandler
5920 longbow drive
boulder, CO 80301
3035306409
MDR Report Key14408722
MDR Text Key291733499
Report Number2936999-2022-00421
Device Sequence Number1
Product Code GXY
UDI-Device Identifier20884521134307
UDI-Public20884521134307
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K093183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/09/2022
Device Model Number186-0106
Device Catalogue Number186-0106
Device Lot Number1209201L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-