| Model Number |
24700 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 8.0mmx60mmx135cm-hybrid sterling balloon catheter was advanced for dilatation.However, during initial inflation at 6 atmospheres for 10 seconds, the balloon ruptured.The device was removed intact from the patient and the procedure was completed with a different sterling balloon catheter.There were no patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 8.0mmx60mmx135cm-hybrid sterling balloon catheter was advanced for dilatation.However, during initial inflation at 6 atmospheres for 10 seconds, the balloon ruptured.The device was removed intact from the patient and the procedure was completed with a different sterling balloon catheter.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 - lot number: updated to 25018380.
|
| |
|