| Model Number |
26012802 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This event will be updated once the investigation is complete.Trends will be evaluated.
|
| |
|
Event or Problem Description
|
|
Allegedly, the femoral head disengaged from the bipolar head unexpectedly.Non-revised products: product id: pha00232, profemur® z femoral stem s 1 cement less/ lot number: 1897645/ qty.1.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to additional information received, it has been indicated that the patient had a fall in the bathroom and the head disengaged unexpectedly from the bipolar cup.This event was caused by the patient fall and the incident is not related to a device failure.Please void this report.
|
| |
|
Event or Problem Description
|
|
Allegedly, the patient underwent bipolar hemiarthroplasty and used the above five implants on (b)(6) 2022.After the surgery, the patient slipped in the bathroom on (b)(6) 2022 and the femoral head disengaged from the bipolar head unexpectedly.The physician operated to replace the neck 8dg a/r (pha01232) with the neck 15dg a/r (pha01242) on (b)(6) 2022.The other four implants (cup, head, liner, stem) didn't have been replaced.
|
| |
|
Search Alerts/Recalls
|