ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There are no other complaints reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
The non-health care professional reported that after intraocular lens implant procedure, the lens found to be dislocated.The lens was exchange with another iol in secondary procedure.Additional information has been requested, however no further information received.
 
Additional Manufacturer Narrative
The lens was returned wrapped in surgical gauzes.Blood and solution was dried on the lens.Both haptics were bent in the gusset and distal area.The optic was cut into pieces, typical of insertion and removal.Qualified associated products were indicated.The product investigation could not identify a root cause for the reported complaint.The issue was not reported at the time of the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14410460
Report Number1119421-2022-01052
Device Sequence Number2619872
Product Code HQL
UDI-Device Identifier00380655093191
UDI-Public00380655093191
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.200
Device Lot Number12501956
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date05/16/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/26/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH CARTRIDGE D; MONARCH INJECTOR; PMMA SINGLEPIECE IOL; PROVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR