• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC MICRO SCREW 1.2MM X 6MM; SCREW, FIXATION, INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC MICRO SCREW 1.2MM X 6MM; SCREW, FIXATION, INTRAOSSEOUS Back to Search Results
Model Number 204-1206
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a facial trauma procedure, the surgeon used the part number 204-1206 screws, and one of the following situations occurred: 1) the screw could not be inserted into bone, and then, the screw was broken or "was not straight" or 2) the screw was inserted into bone, and then, the screw was broken.It was reported predrilling was performed.There were no adverse consequences to the patient; however, it was reported this issue resulted in delay in procedure and fragment of the broken screw remains in the patient.It is unknown how many screws brokes broke/fragments remain in patient.The surgeon used 1.2mm screw with another length to complete the surgery.Information received indicated 14 screws from three different batch/lot numbers were involved.This report is related to the following report numbers for the other screws involved in this event: 2027754-2022-00012, 2027754-2022-00013, 2027754-2022-00014, 2027754-2022-00015, 2027754-2022-00016, 2027754-2022-00017, 2027754-2022-00018, 2027754-2022-00020, 2027754-2022-00021, 2027754-2022-00022, 2027754-2022-00023, 2027754-2022-00024, and 2027754-2022-00025.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive as the device has not been received for evaluation.If the device is returned, a follow up report will be submitted upon completion of the investigation.
 
Additional Manufacturer Narrative
The lot numbers for this event were reported as 1154827, 1157319, and 1160324, so the device history records (dhrs) were reviewed for these lot numbers.The review revealed all inspections passed, and no anomalies were noted.Six (6) screws were returned for evaluation.Additionally, fragments of broken screws were received.The six (6) screws were inspected by quality control (qc) personnel per the inspection plan.The results for each of the six (6) screws were as follows: one screw was observed to be unused and met all specifications.The second and third screws had cruciforms that showed signs of usage/exhibited wear and therefore did not meet specifications.The fourth, fifth, and sixth screws were only able to measured for the length but were unable to be measured for all other features due to significant damage to the cruciform, which is indicative of multiple failed attempts of insertion.A two year review of complaints indicated there have been two (2) complaints for this part family of screws; however, the complaint in this report is the only complaint received for "cannot be inserted into bone and screw was broken/not straight".Review of the instructions for use (ifu) revealed the following guidance is provided: use of an undersized plate or screw in areas of high functional stresses may lead to implant fracture and failure.Use of screws in high density bone may lead to implant fracture or failure upon insertion.Use of excessive torque during insertion of screws may lead to implant failure.All osteomed plates, screws, and instrumentation may be required for each surgery.Failure to use dedicated, unique osteomed instruments for every step of the implantation technique may compromise the integrity of the implanted device, leading to premature device failure and subsequent patient injury.Failed devices may require re-operation and removal.Carefully inspect the osteomed implants prior to use.Inspect the instruments before and after each procedure to assure they are in proper operating condition.Instruments which are faulty, damaged, or suspect should not be used.They should be replaced or sent to osteomed.The root cause of the reported event could not be determined based on the information provided and the investigation performed.Out of the six (6) devices returned, one (1) device which exhibited no signs of wear/usage met specification per the inspection plan while the five (5) remaining devices did not meet specification for one (1) or more features since damage to the cruciform was noted, which is indicative of multiple repeated/failed attempts of insertion; however, due to unknown procedural conditions, the cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO SCREW 1.2MM X 6MM
Common Device Name
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key14411309
Report Number2027754-2022-00019
Device Sequence Number5514252
Product Code DZL
UDI-Device Identifier00845694000482
UDI-Public(01)00845694000482(10)1157319(30)1(11)210414
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K924138
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number204-1206
Device Catalogue Number204-1206
Device Lot Number1157319
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/16/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age44 YR
Patient SexFemale
Patient RaceAsian
-
-