• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SILK BLK 18IN 5-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SILK BLK 18IN 5-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number 640G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The product was tearing and unable to be used.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was there any medical or surgical intervention performed (product removed; re-operation; re-closure; prescription medication)? if so, please specify.- no injuries reported.Can you identify the product code and lot number of the product that was used? ¿ product code 640g.Lots qbbaes, paj177 & pkj281.The following information was requested, but unavailable: when was the product tore and unable to use (in the package, during removal from package, during handling or during use on the patient)? please specify- unknown.Did the event occur during one or multiple patient procedures? unknown.What is the total number of procedures? unknown.Have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).- unknown.If this event occurred in multiple procedures, please provide the following information for each patient event: what is the procedure name? - unknown.What is the procedure date? - unknown.If medication was required, please clarify if it was prescribed by a physician.- none reported.What is the most current patient status? ¿ unknown, no injuries reported.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2022-03759, mw # 2210968-2022-03760.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Date sent to the fda: 6/13/2022 additional information: d9, h3, h6 h6 component code: g07002 no device problem found h3 investigational summary: twelve unopened samples of product code 640g were returned in ethicon for analysis.Upon initial inspection, of the samples, no external damages were observed.Visual inspection was performed to the twelve samples, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during the evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILK BLK 18IN 5-0 S/A P-3 PRM MP
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14414688
Report Number2210968-2022-03758
Device Sequence Number12163994
Product Code GAP
UDI-Device Identifier10705031007369
UDI-Public10705031007369
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number640G
Device Catalogue Number640G
Device Lot NumberQBBAES
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date05/17/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-