| Model Number |
SSLP-250 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iii/iv.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii/iv.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, weight to the spec, white particles on the shell, and crease fold.After autoclave cycle were observed, cloudy color in the gel and voids.Based on the device analysis, the final assessment is: no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side capsular contracture, baker grade iii/iv.Device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|