STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY SYNCHRO SELECT-14 SUPPORT PS 300CM; GUIDE, WIRE, CATHETER, NEUROVASCULATURE Back to Search Results
Model Number SSUP300PRE
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the coil embolization procedure, physician felt resistance in the distal end while operating the subject guidewire.When the subject guidewire was removed to insert the coil it was confirmed that the 4cm tip of the subject guidewire was detached and left inside the patient's anatomy.An attempt was made to retrieve the tip but was difficult so it was embolized with the coils.Patient's anatomy was very meandering.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the coil embolization procedure, physician felt resistance in the distal end while operating the subject guidewire.When the subject guidewire was removed to insert the coil it was confirmed that the 4cm tip of the subject guidewire was detached and left inside the patient's anatomy.An attempt was made to retrieve the tip but was difficult so it was embolized with the coils.Patient's anatomy was very meandering.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date ¿ added.There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.It was reported that resistance was felt when operating the subject guidewire and when it was removed, it was confirmed that 4cm of the guidewire tip had detached.An attempt was made to retrieve it; however, it was found to be difficult and the physician ended up coiling and placing the coil as it was.Additional information provided by the customer indicated that the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared as per the dfu, flush was given inside the microcatheter when the guidewire was inserted, the introducer was used when inserting the guidewire, there was no allegation against the microcatheter used in the case, and the patient's anatomy was very meandering.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint of guidewire distal tip broken/fractured during use, un-retrieved device fragments and guidewire distal end friction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRO SELECT-14 SUPPORT PS 300CM
Common Device Name
GUIDE, WIRE, CATHETER, NEUROVASCULATURE
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14417784
Report Number3012931345-2022-00099
Device Sequence Number5301617
Product Code MOF
UDI-Device Identifier07613327508642
UDI-Public07613327508642
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202522
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/02/2024
Device Model NumberSSUP300PRE
Device Catalogue NumberSSUP300PRE
Device Lot Number0000063351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date05/17/2022
Supplement Report FDA Received Date08/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-