The customer reported that 2 false positive results occurred when using a urine sample on the alere hcg combo test device (serum/urine) prior to an elective colonoscopy procedure.Confirmatory tests with a blood sample were performed with an unknown method, bhcg quant on (b)(6) 2022 was a negative result, 0.2miu/ml.The procedure was to be rescheduled.No adverse evens occurred due to the delay in the colonoscopy procedure.Medications: adderall xr 15 mg extended release capsule take 15 mg by mouth every morning.Dulcolax 5 mg enteric coated tablet please take as directed at bedtime 2 days prior to gi procedure.Catapres 0.1 mg tablet take 1/2 to 1 tablet by mouth at bedtime as needed for sleep.Flexeril 10 mg tablet take 1 tablet by mouth 2 times a day as needed for muscle spasms for up to 7 days.Pristiq 100 mg extended release tablet take by mouth daily.Avapro 300 mg tablet.Lisinopril 5 mg tablet take 5 mg by mouth daily with food.Multivitamin tablet take 1 tablet by mouth daily.Prilosec 40 mg delayed release capsule take 40 mg by mouth daily with food.Golytely solution please take half of prep solution at 9 pm night before gi procedure, and second half of prep 6 hours prior to procedure arrival.Effexor xr 75 mg extended release capsule take 75 mg by mouth daily.
|
|
Product will not be returned.Retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in serum and urine specimens shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning serum or urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in serum or urine specimens should not be used to diagnose pregnancy unless these conditions have been ruled out.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (hama) in the specimen.Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain hama.Such specimens may cause false positive or false negative results.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
|