• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Model Number 7209485
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a meniscus repair, instruments outside patient, the needle was found to be already broken when the package was opened.The procedure was completed with a s+n back up device.There was no delay and no further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was returned to the designated complaint unit for independent evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the device drawings found a stress relief collar is now attached at the connection between the shaft and the hub of the suture capture loop.This design change was implemented as a result of a corrective action initiated to address the reported issue.A visual inspection of the returned device found that it is not in its original packaging.The shafts of the menders are separated from the knobs, which were not returned.The proximal end of one of the devices is twisted and deformed while the other is broken cleanly.There does not appear to be any debris on the devices.The complaint was confirmed, and the root cause was associated with device design.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A corrective action for this failure mode is in place.H11: corrected data d7a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Common Device Name
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14420916
Report Number1219602-2022-00737
Device Sequence Number5474002
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2025
Device Model Number7209485
Device Catalogue Number7209485
Device Lot Number2065162
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date05/17/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-