SMITH & NEPHEW, INC. XL SUTURE CUTTER; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72202588
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an arthroscopy, the tip of the xl suture cutter fell into the patient's body, the pieces remained in the patient and were not removed.The procedure was completed with a s+n back up device.There was a non-significant delay and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the tip of the xl suture cutter fell into the patient's body, the device was removed by the surgeon.The procedure was completed with a s+n back up device.There was a non-significant delay and no further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The device shows wear from use and the product identification markings are significantly degraded.The carbide insert in the distal tip is fractured and there is material missing.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states two intraoperative photos were reviewed and confirm the reported breakage.Of the two images, one arthroscopic photo appears to show the broken tip of the device inside the patient and the second photo appears to show the suture cutter with a missing tip.The patient impact beyond the reported events is not anticipated as the broken device pieces were reportedly removed from the patient and no patient harm has been alleged.The root cause was associated with component failure.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XL SUTURE CUTTER
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14421522
Report Number1219602-2022-00740
Device Sequence Number5527694
Product Code NBH
UDI-Device Identifier03596010652430
UDI-Public03596010652430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202588
Device Catalogue Number72202588
Device Lot Number35077
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer06/29/2022
12/19/2023
Initial Report FDA Received Date05/17/2022
Supplement Report FDA Received Date06/30/2022
12/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-