• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG BIPOLAR CUTTING LOOP; CUTTING LOOP 24/26 BIPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG BIPOLAR CUTTING LOOP; CUTTING LOOP 24/26 BIPOLAR Back to Search Results
Model Number 27040GP140-S
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The proximal tip of the cutting loop was evaluated and confirmed to have been separated from shaft or body of cutting loop.
 
Event or Problem Description
During a turp, the proximal end of the loop broke inside the patient.The broken piece was removed with no harm to patient.The procedure was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR CUTTING LOOP
Common Device Name
CUTTING LOOP 24/26 BIPOLAR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm,
GM  
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo 
MDR Report Key14427180
Report Number9610617-2022-00073
Device Sequence Number20235140
Product Code FAS
UDI-Device Identifier04048551393208
UDI-Public4048551393208
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K061541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number27040GP140-S
Device Catalogue Number27040GP140-S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Initial Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-