| Model Number |
1294-53-236 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised due to the femoral adapter broke off and the morse taper part of the sleeve was stuck in the sleeve and the collar portion was floating with sharp edges in the femur.The metal debris was the reason for revision.We took out all of the femoral components and put in a dfr implant.The surgeon believes the adapter wasn¿t tighten/put together correctly.Unknown if any loosening.There is a reported surgical delay of 1.5 hours.Doi: (b)(6) 2015; dor: (b)(6) 2022; affected side: left.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|